EasyManua.ls Logo

DOD CL96 - Page 12

DOD CL96
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 9 of 44
DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300
DC-MAN-CL96Servicemanual
Innovative Life Safety Systems & Services
Location
DC Voltage
Both 24VDC supplies on pump panel
Greater than 24.6 24.8 VDC
12VDC supply on pump panel
Greater than 12.5VDC
5VDC supply on pump panel
6.1VDC (max)
24VDC in control box for PC
Greater than 24.3VDC
24VDC in control box for I/O
Greater than 24.3VDC
5VDC in control box for flow system
Greater than 5.3VDC
12VDC in Control box
Greater than 12.2VDC
5VDC supply for optic blocks
Greater than 5VDC
*All power supplies can be adjusted with potentiometer
All power supply voltages should be measured from the pump panel (image below) to begin the voltage
verification test and voltages should be adjusted with the potentiometer if outside of the factory
specification.
In the Control Box there is a row of terminal blocks mounted on the rear panel (see image below). The
supply voltage for the PC and monitor can be measured at the yellow terminal block. The 12VDC supply
can be measured at the orange terminal block and the 5VDC for the flow system can be measured at the
red terminal block. The DC voltage verification should be done at the yellow terminal blocks mounted on
24V
24V
12V
5V