EasyManua.ls Logo

DOD CL96 - Page 16

DOD CL96
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 13 of 44
DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300
DC-MAN-CL96Servicemanual
Innovative Life Safety Systems & Services
96116
Low Flow - Verify Tray Insertion
The flow is low for multiple points on an
analyzer.
1. Verify analyzer is
fully inserted.
2. Make sure all
points are flow
balanced. 3.
Refer to Flow Path
verification procedure
96117
Modbus Comm. Failure
Interruption during Modbus
communication process.
1. Power cycle unit
2. Contact DOD
Technologies, Inc.
96118
Optic Block Comm. Failure
Communication interruption between
optic block and PC.
1. Refer to Optic
Troubleshooting
Procedure.
96119
Flow Block Comm. Failure
Communication interruption between
Flow board and PC.
1. Refer to Flow
Comm. Fault
Procedure.
96121
Gate Open Fault
Micro switch failed to open during a
tape advance.
1. Refer to Gate Fault
Procedure.
96122
Gate Close Fault
Micro switch failed to close during a
tape advance.
1. Refer to Gate Fault
Procedure.
96123
Setting DAC to preset values
CL96 was unable to set DAC values back
to previously retained values.
1. Power cycle
analyzer
96130
System Fault
System hardware issue.
1. Contact DOD
96131
Block Below Minimum Voltage
Optic block voltage dropped below the
required voltage of 5.1 V
1. Verify 5V supply is
turned up to 6.1 VDC
2. Verify all optic
blocks have at least
5.4 VDC
3. See Optic
Troubleshooting
Procedure
96133
Output Module Failure or Not
Installed
Expected I/O module cannot be located
by the PC.
1. Refer to I/O
Verification
Procedure.
96138
Retentive Memory Error, History
and/or Tape counters may be
corrupt.
Unable to load previous historical
events after power cycle.
1.Power cycle CL96
96140
Idle Timeout (System)
CL96 was left out of analysis for a time
frame that overlapped the time out
setting.
1. Check to see if a
fault caused the CL96
to leave analysis
2. If not put the CL96
back into analysis