EasyManua.ls Logo

DOD CL96 - Optic System Troubleshooting

DOD CL96
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 17 of 44
DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300
DC-MAN-CL96Servicemanual
Innovative Life Safety Systems & Services
The screen above should be used to locate the point(s) that need to be cleaned. Press the Calibrate
button to allow the optics to adjust. Locate any point that has a DAC value above 140. Once the points
that need to be cleaned are located gather the following tools:
1. # 2 Phillips screw driver.
2. Cotton swabs
3. A form of Alcohol (methanol, ethanol)
The top of each analyzer will have a point map; use this map to locate the optic block that needs to be
removed. Once the block is located open the related gate by pressing the open gate button and remove
the ChemLogic cassette out of the way. Next, remove the optic block cover by unscrewing out the two
phillips screws (see image below).