DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300
If a 0 value is found press the Set Offsets button for the related analyzer. If this does not resolve the
issue check the flow system fuse light (See image below). If the red light is on, the fuse is blown and
must be replaced.
Next remove the rear panel and locate the flow system interconnect board (see image below). Verify
that there is greater than 5.3V at the red (+) and black (-) wires. Also, verify that all interconnect board
connections are secured.