Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 109
Getting started
WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
2
flowmeter. Leave the flow control valve completely closed.
2. Open the flow control valve (1) and set the desired O
2 flow. If necessary, press
the O
2+ key (3) to quickly fill the breathing bag.
3. Monitor the flow on the total flow tube (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the Standby page is displayed, start the therapy, see page 143.
6.7.3 Electronically controlled gas mixer
1. Adjust the APL valve.
2. Set the O
2 switch (3) to the Add. O2 position.
WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
2
flowmeter. Leave the flow control valve completely closed.
3. Open the flow control valve (2) and set the desired O
2 flow. If necessary, press
the O
2+ key (4) to quickly fill the breathing bag.
4. Monitor the set flow on the O
2 flowmeter (1).
5. Set the anesthetic gas concentration on the vaporizer.
6. Ventilate the patient manually.
7. Turn on the device.
8. As soon as the Standby page is displayed, start the therapy, see page 143.
9. Set the O
2 switch (3) to Aux. O2.
35041