202 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Configuration
11.3 Device settings
Some device functions are available as an option and consequently are only
available on appropriately equipped devices.
11.3.1 Factory defaults
Dräger delivers the device with factory defaults that are used when starting the
device for the first time. Service personnel can reset the device to the factory
defaults.
11.3.2 Start settings
Start settings take effect after every restart of the device or when starting a new
case (touching one of the New adult, New pediatric or New neonate buttons). The
start settings can be adjusted after the user password is entered.
If required, the device can be delivered with start settings that may differ from the
factory defaults.
11.3.3 User-specific settings
User-specific settings can be adjusted by the user without a user password. The
settings take effect immediately but are discarded at the latest after a device restart.
11.4 Setting the date and time
The device can adopt the time from a network or from a device connected via
MEDIBUS. The time synchronization takes place shortly after switch-on and at
regular intervals thereafter.
If time synchronization is not set, the time can be changed manually in 2 ways:
● Touch the field (1).
● Set the date and time in the system setup.
The source for the time synchronization can also be set in the system setup. For
further information, see the following chapter: "Vertical tab "General"", page 217.
36019