Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 189
Ending operation
CO2 absorber
● Check the soda lime for discoloration and replace it if necessary, see page 93.
Breathing hoses and filters
1. Replace the hoses and filters according to infection prevention policy of the
health-care facility.
2. Connect a suitable breathing circuit and filters, see page 85.
8.15.3 Checking the device
Prerequisite: The device is in standby mode.
1. Perform the leakage test, see page 112.
2. Flush the breathing system if necessary.
9 Ending operation
9.1 At the end of the OR day
Dräger recommends shutting down the device during longer periods of non-use
such as overnight or on weekends. This can lower power consumption and prolong
the life span of the medical device without negatively influencing device availability.
Before shutting down, the device can optionally be prepared for automatic switch-
on including an automatic system test.
9.1.1 End of operation
1. Make sure that all flow control valves are closed.
2. Press the key.
3. Follow the instructions on the screen.
9.1.2 Programming the system test (Auto On)
Using the Auto On function, the device can be programmed so that it performs an
automatic system test and is tested at a specific time.
During the initialization, the Auto On function checks the components that
frequently cause irregularities, e.g., high leakage, insufficient gas supply. After that,
the device remains in Standby mode until the automatic system test, or it is shut
down.
1. Press the key.
2. Touch the Auto On preparation button.
3. Prepare the device in accordance with the instructions on the screen. For details
on the test steps, see page 125.
4. Set the desired day and time for the device to be ready for operation.
5. Select whether the device should remain in Standby mode or be shut down.
6. Touch the Start button.
7. If necessary, observe any messages that occur.