262 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Troubleshooting
! 255 O2 concentration implausi-
bly high
The O2 sensor calibration is
invalid.
Use substitute O2 monitoring.
When the case has been
completed, perform a sys-
tem test.
! 220 O
2 cylinder almost empty The cylinder is almost empty. Replace the cylinder. Use the
central supply.
!!! 210 O
2 cylinder empty The cylinder is empty or
closed.
Replace the cylinder. Use the
central supply.
Use "ALARM RESET" to
downgrade alarm priority.
! 255 O
2 cylinder empty The cylinder is empty or
closed.
Replace the cylinder. Use the
central supply.
! 190 O
2 cylinder sensor? Cylinder pressure sensor is
not connected.
Check if cylinder pressure
sensor is connected.
Use "ALARM RESET" to
reset alarm.
!! 150 O
2 FG flow measurement
failed
The measurement system for
the O
2 fresh-gas flow has
failed.
Only use O
2 as fresh gas.
Use the total flow tube to
check the fresh-gas flow.
Adjust the fresh-gas flow so
that it equals or exceeds the
minute volume.
!! 255 O
2 measurement not
available
The ambient air used for cali-
brating the sensor was
impure.
Position the device in an
environment with clean ambi-
ent air.
Wait for the automatic cali-
bration.
There is electromagnetic
interference.
Check for electromagnetic
radiation in the vicinity.
Ambient temperature is too
high.
Check ambient conditions.
Use "ALARM RESET" to
downgrade alarm priority.
! 255 O
2 measurement tempo-
rarily inaccurate
The sensor has not yet
warmed up.
Wait for the automatic cali-
bration.
The ambient air used for cali-
brating the sensor was
impure.
Position the device in an
environment with clean ambi-
ent air.
Wait for the automatic cali-
bration.
!! 255 O
2 sensor failure The O2 sensor in the patient-
gas measurement module is
faulty.
Use substitute O2 monitoring.
Use "ALARM RESET" to
downgrade alarm priority.
Perform the system test.
Call Dräger.
Priority Alarm Cause Remedy

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals