Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 217
Configuration
11.5.6 System setup > System
11.5.6.1 Vertical tab "General"
Headline/
Parameter
Setting range
Description
Language List of available languages
English (United States)
Selects the language.
A flag symbol identifies the tabs that
lead to the page with the language set-
tings.
Height above sea
level [m]; [ft]
[m]: 0 up to 4000
[ft]: 0 up to 13120
Sets the height above sea level.
Time source MEDIBUS 1; MEDIBUS 2; NTP server;
None
Selects the source for the time synchro-
nization.
Prerequisite: The connected device
supports this function.
Date and time day; month; year
hour; minute
Sets the date and time. The change is
applied on leaving the "General" verti-
cal tab.
Automatic switch
to daylight sav-
ings time
On; Off Activates or deactivates the automatic
changeover to daylight saving time.
OR working hours Hour : Minute to Hour : Minute
6:30 to 18:30
Sets the working hours of the operating
room.
During this time, the gas measurement
is kept in a pre-warmed and calibrated
state so measured values are available
after only a short waiting period. How-
ever, this decreases the life span of the
patient-gas measurement module.
Device name Device name (up to 16 alphanumeric
characters)
A3XX
Changes the device name in order,
e.g., to enter the operating location.
User password Change password Changes the user password.
Reset all pages to Factory defaults Resets all the settings on all pages in
the System setup dialog to the factory
defaults.

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals