216 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Configuration
11.5.5 System setup > Licenses/Options
11.5.5.1 Vertical tab "Licenses/Options"
Auto Corrects the measured flow values
automatically when an anesthetic gas
monitor is connected.
Air Corrects the measured flow values
when Air is used.
N
2O Corrects the measured flow values
when N
2O is used.
Enable pause
mode
On; Off Defines whether the Pause mode is
available.
On: Pause mode is activated.
Off: Pause is deactivated.
For devices with integrated patient-gas
measurement module, Monitoring
mode is available as an alternative.
Headline/
Parameter
Setting range
Description
Licenses for software options Overview of available and active
software options.
Activating software options.
Observe the information in the following
section: "Activating software
options", page 224
Gas supply
Disable N
2OOn; Off A device with electronic gas mixer and
connectors for nitrous oxide can be
configured so that nitrous oxide is no
longer displayed and cannot be
selected as a carrier gas.
Prerequisite:
The current carrier gas is Air.
Nitrous oxide is not connected or
available.
Restart the device if necessary to
implement the settings.
On: Nitrous oxide cannot be delivered.
Off: Nitrous oxide can be delivered.
Headline/
Parameter
Setting range
Description

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals