EasyManua.ls Logo

Dräger Atlan A100

Dräger Atlan A100
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
216 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Configuration
11.5.5 System setup > Licenses/Options
11.5.5.1 Vertical tab "Licenses/Options"
Auto Corrects the measured flow values
automatically when an anesthetic gas
monitor is connected.
Air Corrects the measured flow values
when Air is used.
N
2O Corrects the measured flow values
when N
2O is used.
Enable pause
mode
On; Off Defines whether the Pause mode is
available.
On: Pause mode is activated.
Off: Pause is deactivated.
For devices with integrated patient-gas
measurement module, Monitoring
mode is available as an alternative.
Headline/
Parameter
Setting range
Description
Licenses for software options Overview of available and active
software options.
Activating software options.
Observe the information in the following
section: "Activating software
options", page 224
Gas supply
Disable N
2OOn; Off A device with electronic gas mixer and
connectors for nitrous oxide can be
configured so that nitrous oxide is no
longer displayed and cannot be
selected as a carrier gas.
Prerequisite:
The current carrier gas is Air.
Nitrous oxide is not connected or
available.
Restart the device if necessary to
implement the settings.
On: Nitrous oxide cannot be delivered.
Off: Nitrous oxide can be delivered.
Headline/
Parameter
Setting range
Description

Table of Contents

Other manuals for Dräger Atlan A100

Related product manuals