Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 215
Configuration
I:E changes with
RR in SIMV
On; Off Determines whether the change of RR
automatically causes a change of I:E. In
this way, Ti remains constant.
Applies to the following ventilation
modes:
–PC - SIMV
– PC - SIMV / PS
–VC - SIMV
– VC - SIMV / PS
– VC - SIMV / AutoFlow
– VC - SIMV / PS / AutoFlow
Prerequisite: Inspiratory-time control is
set to I:E.
Breathing system
warmer
On; Off Switches the breathing system warmer
on or off.
Auto Wake-up
(opens start dia-
log upon respira-
tory activity)
On; Off Determines whether the Start dialog
opens automatically when ventilation
activity is detected (e.g., as a result of
repeated squeezing of the breathing
bag).
Anesthetic gas for
flow correction
Auto; Des; Sev;
Other; None
Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
2 measurement.
Auto Corrects the measured flow values
automatically when an anesthetic gas
monitor is connected (see "Flow correc-
tion", page 149).
None Deactivates the correction.
Sev 0 to 6.5
2.1
Corrects the measured flow values
when sevoflurane is used.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Des 0 to 18
6
Corrects the measured flow values
when desflurane is used.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Other Deactivates the correction.
Carrier gas for
flow correction
Auto; Air; N
2O Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
2 measurement and mechanically
controlled gas mixer with flow tubes.
Headline/
Parameter
Setting range
Description