214 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Configuration
11.5.4.6 Vertical tab "Patient"
11.5.4.7 Vertical tab "General"
Headline/
Parameter
Setting range
Description
Default selection
for "Start" dialog
Continue case;
New adult; New pediatric; New neonate
Specifies which button is preselected
the first time the Start dialog is opened,
and after changes to the system setup.
Weight
[kg]
[lb]
Off;
30 to 300
Off
Off; 67 to
661
176
Off;
5 to 50
Off
Off;12 to 110
55
0.4 to 10
3.0
0.9 to 22
6.6
Specifies the following starting values:
– Body weight
– Body height
– Patient age
If the setting is configured to Off, the
corresponding therapy control will not
be displayed in the Start dialog.
Height
[cm]
[in]
120 to 300
185
48 to 118
73
50 to 300
100
20 to 118
39
Off; 20 to 80
Off
Off; 8 to 31
Off
Age [years, Neo:
months]
12 to 130
32
0 to 16
8
0 to 24
6
Headline/
Parameter
Setting range
Description
Pinsp changes
with PEEP
On; Off Determines whether changing PEEP
automatically also changes Pinsp, so
that the difference between PEEP and
Pinsp remains constant.
Inspiratory-time
control
Ti; I:E Determines whether the ventilation set-
tings are based on Ti or on the I:E ratio.
Ti changes with
RR in CMV
On; Off Determines whether the change of RR
automatically causes a change of Ti. In
this way, the I:E ratio remains constant.
Applies to the following ventilation
modes:
–PC - CMV
–VC - CMV
– VC - CMV / AutoFlow
Prerequisite: Inspiratory-time control is
set to Ti.