Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 213
Configuration
11.5.4.5 Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Pressure
decrease per step
when PEEP ≤ 15
[mbar]; [hPa];
[cmH
2O]
1 to 10
2
1 to 10
2
1 to 10
2
For PEEP ≤15: Specifies the pressure
by which PEEP and Pinsp will be
reduced in steps.
Headline/
Parameter
Setting range
Description
Start settings for fresh gas Selects the start settings for the fresh-
gas delivery.
FG O
2
[%]
21 to 100
100
21 to 100
100
21 to 100
100
Sets the O
2 concentration in the fresh
gas.
FG flow
[L/min]
0.20 to
15.00
2.00
0.20 to
15.00
2.00
0.20 to
15.00
2.00
Sets the fresh-gas flow.
Minimal O
2 flow
(carrier gas: Air)
[mL/min]
Off; 50 to
300
200
Off; 50 to
300
200
Off; 50 to
300
200
Sets the minimal O2 flow that is deliv-
ered when Air is used as carrier gas.
Do not set the value too low, but suit-
able for the patient category. The rec-
ommendation is, e.g., 200 for adults,
100 for pediatric patients, and 50 for
neonates.
Minimal O
2 flow
(carrier gas: N
2O)
[mL/min]
50 to 300
200
50 to 300
200
50 to 300
200
Sets the minimum O
2 flow that is deliv-
ered when N
2O is used as carrier gas.
Do not set this value too small; recom-
mended is, e.g., 200 for adults, 100 for
pediatric patients, and 50 for neonates.
Carrier gas Air; N
2O Air; N2O Air; N2O Sets the carrier gas.