118 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Tests
2. Touch the Start button.
a. Wait until the leakage value is stable. Note down the value as the leakage of
the inspiratory hose.
If the leakage value exceeds 500 mL/min, use another breathing circuit.
3. Touch the OK button to switch back to the display of available tests.
4. To test the outer hose, connect the breathing circuit as follows:
a. Connect the hose (7) to the inspiratory port (8).
b. Remove the coaxial test adapter (9) from the expiratory port (10).
c. Connect the elbow (11) to the hose (12).
d. Plug the hose (12) with the elbow connected on to the circuit plug (13).
e. Connect the hose (14) to the expiratory port (4).
5. Touch the Leakage test or System test button to start an automatic test.
The determined leakage value is displayed as Leakage (mech. vent.). Note
down the value as the leakage of the expiratory hose.
39555
9
11
14
12
10
7
8
13