Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 133
Tests
7.5.5.3 Manual resuscitator
1. Make sure there is a manual resuscitator on the device.
2. Check the functional integrity of the manual resuscitator.
7.5.5.4 Anesthetic gas receiving system
With active anesthetic gas scavenging:
Have the flow for the anesthetic gas scavenging system set so that the flow
indicator (1) floats within range (2) ("normal range").
If the flow indicator (1) is floating within range (3) ("restricted range"), certain
fresh-gas flows should not be exceeded, see "Anesthetic gas receiving system"
in chapter "Technical data". Contamination of the ambient air can be prevented
by limiting the fresh-gas flow.
3512135125
2
3
1