184 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Operation
2. Touch the Therapy tab (1).
3. Touch the appropriate button (2).
When switching to standby mode, the heater is reset to the value configured in the
system setup.
CAUTION
Risk due to faulty or switched-off breathing system warmer
Without the breathing system warmer, increased condensation may occur in the
breathing system. This may also impair the flow measurement and the metering
accuracy of the set tidal volume.
Increase the fresh-gas flow if necessary.
Regularly remove the condensed water from hoses, water traps, and the
breathing system.
Have service personnel repair the faulty breathing system warmer.
8.13.2 Creating additional logbook entries
The following events can generate a logbook entry with measured values for the
parameters etCO
2, MV, Pmean, PIP, Pplat, PEEP, FiO2, expiratory concentration
of the primary anesthetic gas, and etN
2O:
Settable interval
Alarms with high or medium priorities
1. Open the System setup dialog.