Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 199
Alarms
Gas measurement alarms with inspiratory O2 measurement
Ventilation alarms
O
2 measurement not
available
N
2O measurement not
available
Agent measurement
not available
O
2 measurement tem-
porarily inaccurate
CO
2 sensor accuracy
low
Agent measurement
temporarily inaccurate
N
2O measurement
temporarily inaccurate
Measured gas concen-
trations temporarily
inaccurate
>20 s --- ---
Alarm Priority Priority Priority
Low Medium High
FiO
2 low --- After at least 5 s
FiO
2 high --- After at least 5 s
O
2 measurement not
available
O
2 measurement tem-
porarily inaccurate
O
2 sensor not ready
>20 s --- ---
Alarm Priority Priority Priority
Low Medium High
Apnea (no flow) --- At the latest after 20 s
(for RR 6)
or
At the latest after 35 s
(for RR <6 or in Manual /
Spontaneous mode)
15 s later
(for RR 6)
or
30 s later
(for RR <6 or in Manual /
Spontaneous mode)
Apnea (no pressure) --- At the latest after 20 s
(for RR 6)
or
At the latest after 35 s
(for RR <6)
15 s later
(for RR 6)
or
30 s later
(for RR <6)
Apnea Ventilation At the latest after 40 s in CPAP / PSV mode with
active RRmin.
(configurable alarm priority, see page 208)
---
Alarm Priority Priority Priority
Low Medium High