EasyManuals Logo

Dräger Atlan A100 User Manual

Dräger Atlan A100
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 199
Alarms
Gas measurement alarms with inspiratory O2 measurement
Ventilation alarms
O
2 measurement not
available
N
2O measurement not
available
Agent measurement
not available
O
2 measurement tem-
porarily inaccurate
CO
2 sensor accuracy
low
Agent measurement
temporarily inaccurate
N
2O measurement
temporarily inaccurate
Measured gas concen-
trations temporarily
inaccurate
>20 s --- ---
Alarm Priority Priority Priority
Low Medium High
FiO
2 low --- After at least 5 s
FiO
2 high --- After at least 5 s
O
2 measurement not
available
O
2 measurement tem-
porarily inaccurate
O
2 sensor not ready
>20 s --- ---
Alarm Priority Priority Priority
Low Medium High
Apnea (no flow) --- At the latest after 20 s
(for RR 6)
or
At the latest after 35 s
(for RR <6 or in Manual /
Spontaneous mode)
15 s later
(for RR 6)
or
30 s later
(for RR <6 or in Manual /
Spontaneous mode)
Apnea (no pressure) --- At the latest after 20 s
(for RR 6)
or
At the latest after 35 s
(for RR <6)
15 s later
(for RR 6)
or
30 s later
(for RR <6)
Apnea Ventilation At the latest after 40 s in CPAP / PSV mode with
active RRmin.
(configurable alarm priority, see page 208)
---
Alarm Priority Priority Priority
Low Medium High

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals