Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 209
Configuration
11.5.3.5 Vertical tab "Config. 2"
11.5.3.6 Vertical tab "Config. 3"
Headline/
Parameter
Setting range
Description
Deactivate the alarm limit in the following modes: Man/Spon,
Pause, CPAP/PSV with ∆Psupp < 5, ext. FGO
1)
Deactivate the alarm limit in the following modes: Man/Spon,
CPAP/PSV with ∆Psupp < 5, ext. FGO, Monitoring
2)
1) Pause must be activated in the system setup.
2) Only for devices with integrated PGM. Pause must be deactivated in the system setup.
FiO2 high Yes; No Yes; No Yes; No Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only when switch-
ing to a ventilation mode with lower or
no breathing support (see page 310).
The alarm behavior at the start of the
therapy is defined by the configuration
in the vertical Limits tab.
MV low Yes; No
MV high Yes; No
xMAC low Yes; No
etCO
2 low Yes; No
etCO
2 high Yes; No
inCO
2 high Yes; No
inAgent low Yes; No
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
2 high Yes; No Specifies the alarm behavior in the
CBM mode.
MV low Yes; No
MV high Yes; No
inAgent low Yes; No
Headline/
Parameter
Setting range
Description
Alarm limits for "Cylinder almost empty"
O
2 [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
Specifies the alarm limits for the supply
pressure of connected gas cylinders.
Air [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
N
2O [bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290