248 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Troubleshooting
!!! 220 Apnea No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
!! 255 Apnea No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
!!! 220 Apnea (no CO
2) No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Sample line is not con-
nected.
Connect sample line to
breathing circuit.
!! 255 Apnea (no CO
2) No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Sample line is not con-
nected.
Connect sample line to
breathing circuit.
!!! 220 Apnea (no flow) No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
!! 255 Apnea (no flow) No breathing or ventilation. Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
Priority Alarm Cause Remedy