Technical data
15.3 Fresh-gas delivery
For storage longer than 12 months –15 °C to 25 °C
(5 °F to 77 °F)
Maximum storage duration without
recharging
180 days
Ambient pressure 500 to 1060 hPa
(7.3 to 15.3 psi)
Relative humidity 10 to 95 %, without condensation
CO
2 concentration Not relevant
Information The permissible ambient conditions depend on the
accessories used. Follow the corresponding instruc-
tions for use.
Gas mixer The device is equipped with either an electronically
controlled gas mixer or a mechanically controlled gas
mixer. The mechanically controlled gas mixer is avail-
able with electronic flow measurement or with flow
tubes.
Standardization All data are standardized to STPD conditions.
O
2 flush 25 to 75 L/min
At 2.7 to 6.9 kPa x 100 [bar] (39 to 100 psi; 0.27 to
0.69 MPa) supply pressure
O
2 flow with integrated flowmeter for O2
insufflation
Setting range Off to 20 L/min
At a low supply pressure of 2.7 kPa x 100 [bar] (39 psi
or 0.27 MPa), the maximum O
2 flow can be limited to
10 L/min.
Accuracy ±10 % of the set value for flows >2.0 L/min
Resolution of displayed value 1 L/min (in the range from 2 to 10 L/min)
5 L/min (above 10 L/min)
O
2 flow with external flowmeter for O2
insufflation
Setting range Off to 15 L/min
At a low supply pressure of 2.7 kPa x 100 [bar] (39 psi
or 0.27 MPa), the maximum O
2 flow can be limited to
10 L/min.
Accuracy ±10 % of the set value
Resolution of displayed value 1 L/min (below 10 L/min)
5 L/min (above 10 L/min)