Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 315
Principles of operation
16.2.5.4 VC - SIMV
Volume-controlled
Pressure-limited
Time-controlled
Machine-triggered or patient-triggered
Constant inspiratory flow
Synchronized inspiration
In VC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing of the patient. If inspiratory effort
by the patient is detected during the inspiratory trigger window, a patient-triggered
breath will be initiated.
16.2.5.5 VC - SIMV / PS
This mode is similar to VC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with Psupp
when outside the trigger window.
16.2.5.6 VC - AF
For volume-controlled mandatory breaths with AutoFlow, the set tidal volume VT is
applied with the lowest required pressure. The pressure patterns and flow patterns
of the mechanical inspiratory breaths correspond to those of pressure-controlled
ventilation.
36290
7ULJJHUZLQGRZ
LQVS
7LPH
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
H[S
Flow
Paw
Pmax
PEEP
Ti
1/RR
Psupp
Insp term
Trigger
Slope
Slope
Trigger

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals