Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 71
Assembly and preparation
6. Insert the new O2 sensor cell (6) into the O2 sensor.
After the O2 sensor cell has been inserted, the initialization phase of the O2
sensor takes place. The initialization lasts 30 minutes. The 30 minutes start when
the mains power supply is established.
4.3.2 Assembling and inserting the O2 sensor
1. Place the sensor cap (1) on the O2 sensor.
2. Turn the knurled nut (2) about 90° clockwise. Turn it until the palpable resistance
is overcome and a click is heard. The sensor cap is now fitted.
39637
6
3717837179