Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 89
Assembly and preparation
1. Connect the breathing bag to the breathing bag hose using the connection
nozzle. Attach the breathing bag hose to the bag elbow.
WARNING
Risk due to pinched breathing bag
If the breathing bag is pinched, excessive airway pressures or a lack of fresh gas
may result.
When attaching, ensure the following:
The breathing bag is not pinched.
The breathing bag can inflate freely.
2. Lay the breathing bag hose in the holder (1) in a loop so that the bag hangs on
the left of the holder.
Make sure that the breathing bag is not impeded by breathing hoses or cables
when inflating.
4.8.3.2 Attaching the breathing bag to the breathing bag arm
Attach the breathing bag to the elbow.
4.8.4 Observing resistance and compliance
Accessories such as filters influence the dead space, compliance, and resistance of
the breathing circuit.
In addition, changes in resistance and compliance arise over time as a result of
moisture in the breathing gas or residues of secretions.
3713737138

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals