Setting the ventilation parameters
Prerequisite: The Basic settings page must be open in
the Ventilator Settings dialog window.
1 Touch the relevant therapy control, e. g., (A).
2 Turn the rotary knob to set the value.
3 Press rotary knob to confirm.
Additional ventilation parameters derived from the ven-
tilation parameter are calculated by Evita XL and dis-
played in the setting assistance field (B).
When the limit set for the parameter has been reached,
Evita XL displays a message.
Setting additional functions
Prerequisite: The Ventilator Settings dialog window
must be open.
1 Touch Add. settings tab (A).
The additional settings of the active or selected ventila-
tion mode are displayed.
2 Touch the tab for the relevant additional setting,
e. g., (B).
The page for setting the associated parameter is dis-
played.
3 Touch therapy control (C).
4 Set the value using the rotary knob and confirm.
To switch on an additional setting:
5 Touch (D) button and press rotary knob to confirm.
To switch off an additional setting:
6 Touch (E) button and press rotary knob to confirm.
The ventilation modes can be combined with additional
settings to optimize ventilation.