22 Pocket Guide Evita XL
Operation
Standby Plus mode (SW 7.0 or higher)
If the device is switched from
Mask Ventilation ap-
plication mode to Standby, Evita XL switches to
Standby Plus.
In Standby Plus mode Evita XL starts ventilation as
soon as it detects the first breath of the patient.
On-screen displays:
Standby (A) is displayed in the header bar of the
screen. Standby Plus (B) is also displayed on the
screen.
When the mask is put back on, Evita XL detects the pa-
tient's inspiration and continues ventilation with the pre-
vious settings. The
Mask Ventilation application
mode is active again.
Flow reduction during disconnection
When the mask is removed, Evita XL reduces the inspi-
ratory flow in the
Adult and Ped. patient cate-
gories. The function can be configured.
Extended leakage compensation
With the NIV Plus option, Evita XL performs leakage
compensation during pressure-controlled ventilation in
the
–
Adult patient category, up to 180 L/min
–
Ped. patient category, up to 60 L/min
–
Neo. patient category, up to 30 L/min
Standby mode (SW 6.12 or higher)
Switch to Standby mode or Standby Plus mode for the
following actions:
– To keep Evita XL ready for operation while the pati-
ent is absent
– To switch between ventilation and O
2 Therapy
– To change application mode
– To change between patient categories
– To perform the Device Check and Airtight Check.
Selecting Standby mode
z Press and hold the
Start/Standby key for
3 seconds.
Evita XL is in Standby mode.
or
z Press
Start/Standby key.
Evita XL opens the Start / Standby dialog window.
1 Touch Standby button (A).
2 Press rotary knob to confirm.
3 Touch Alarm Reset button (B) in the header bar.
4 Press rotary knob to confirm.
Evita XL is in Standby mode or Standby Plus mode (only
with NIV Plus option).