Pocket Guide Evita XL 31
Alarm messages
Standby activated !!! 255 Evita XL has been switched to
Standby.
Acknowledge Standby with
'Alarm Reset' button and confirm.
Tidal volume high !!! 165 The upper alarm limit of the ap-
plied inspiratory tidal volume V
Ti
has been exceeded during three
consecutive ventilator breaths.
Check condition of patient. Check
pattern of ventilation. Adjust
alarm limit if necessary.
Leak or disconnection. Check that patient circuit and
connections are leakproof.
Vol. not const., pressure
limited !!
220 Due to pressure limitation or
time limit, the set tidal volume
V
T has not been delivered.
Prolong inspiratory time 'T
insp', in-
crease inspiratory flow 'Flow', in-
crease pressure limitation 'P
max.
Touch the 'Alarm Reset' button
and confirm to suppress the visu-
al and audible alarms until the
cause of the alarm is remedied.
Message Priority Cause Remedy