Overview
This chapter describes the necessary steps to start
monitoring a patient on the IACS.
Specifically, this chapter describes how to:
Turn the IACS on/off
Admit/discharge a patient on the Cockpit
Change the patient category
Turning the IACS on/off
Before monitoring a patient on the IACS, the
Cockpit and the M540 must be turned on. The
following steps assume that the M540 has been
docked in the M500 (for information see,
"Docking/undocking the M540" on page 93).
To turn the Cockpit on
Press the on/off key (B) of the Cockpit.
The LEDs (A) and the on/off key light up green.
The Cockpit emits a power-up tone and per-
forms a selftest. After a brief moment, the
Dräger startup screen appears before the Cock-
pit main screen appears.
To turn the M540 on
Press the on/off key (C) of the M540.
The M540 emits a high-pitched tone followed by
two power-up tones, performs a selftest, and
displays the New patient prompt. Select
Discharge to delete the previous patient data
or Cancel to continue monitoring the patient
and append the new data to the previous data
set. The main screen appears.
The acoustic alarm signals are paused for
2 minutes.