Impedance respiratory rate (RRi)
248 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Accessing the respiration settings
Select the ventilation parameter field to select
the Resp. page directly.
or
1 Select Sensor parameters... from the main
menu bar.
2 Select the Resp. tab to access the Resp. page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog: symbol and the display filter
button.
3 Select the Settings 1 and Settings 2 tabs.
The top portion of the Settings 1 page contains
the Auto set and Alarm buttons for configuring
the alarm functions. For detailed alarm setup
information, see "Configuring the alarm settings
for a patient" on page 125.
Ventilation parameter setup functions
All respiration setup functions take place in the
Resp. page.
Selection Available settings Description
Settings 1 page
Resp. lead I, II (default) Selects the lead for respiration
monitoring.
Relearn None Initiates a relearning of the
respiration signal.
Mode Auto (default), Manual (see page 247 for
more details).
Determines the processing
mode for the breath-related
impedance change.
Size [%] 10% to 100% (in 10% increments) – default:
50%
Adjusts the waveform size
and/or breath detection
threshold, according to the
selected respiration setting.
Auto mode – Waveform
size only, without affecting
the breath-detection
threshold.
Manual mode – Waveform
size and breath-detection
threshold.