Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 361
Scio Monitoring
Using the Scio dialog
Setup for Scio parameters takes place in the Scio
dialog. This dialog contains the following tabs:
–CO2
–O
2
Agent settings
Agent alarms
–Show all
To access the Scio settings
Scio settings are available through the CO2, O2, or
Agent parameter fields.
If the CO2, O2, or Agent parameter field displays on
the Cockpit:
1 Touch the CO
2, O2, or Agent parameter field.
2 Touch the Scio tab.
Or, if the CO
2, O2, or Agent parameter field is not
displayed:
Select Sensor parameters... from the main
menu bar > Scio tab to access the Scio page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog: symbol and the display filter
button.
CO2 alarm limits
The CO2 tab displays fields on the upper left side of
the window for etCO2, inCO2, and RRc alarm limit
settings.
etCO2 and RRc allow adjustment of the upper and
lower alarm limits, use of the Archive feature, and
use of Auto set. For inCO
2, only the upper limit and
Archive can be used.