Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 551
Troubleshooting
Cardiac Output (C.O.)
Priority Message Parameter field
Problem Solution
!! Tblood > (alarm limit)
Tblood < (alarm limit)
Parameter value The blood
temperature is
outside the alarm
limits because of:
a physiological
condition
inappropriate
alarm limits
a faulty sensor
Check the patient
and treat if
necessary.
Change the alarm
limits.
!! %0 out of range
high
1)
%0 out of range
low
1)
Parameter value The blood
temperature is
outside the
measurement
range because of a
faulty sensor.
Check equipment and
replace if necessary.
! C.O. Catheter Fault -
Bad Ref.
2)
Parameter value The C.O. blood
thermistor
calibration resistor
does not meet the
specified
tolerance.
Check the
catheter and
replace if
necessary.
Contact
DrägerService
®
.
! C.O. Pod Fault - Bad
Ref.
2)
Parameter value The C.O.
reference values
do not meet the
specified
tolerances.
Remove and
reconnect the
pod.
Repeat the
measurement.
Replace the pod
and contact
DrägerService if
the message
persists.
1)
%0 is a placeholder for the parameter label C.O.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.