Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 573
Maintenance
Maintenance Restart
To maintain optimal performance, the Cockpit must
be restarted on a regular basis. When specific IACS
restart thresholds are reached, the Cockpit pro-
vides a message to the user as described in the fol-
lowing table.
Restart
Threshold
Mode(s) of
operation
Message content Message
duration
Options Result(s)
Voluntary Discharge Restart is required for
scheduled
maintenance
30 seconds Select Restart Restarts IACS
immediately
Records
"Restart is
required"
message to
alarm history
Select Cancel Delays the next
restart message
until a semi-
critical or critical
threshold is
reached
No action taken
during message
duration
Restarts IACS
Records
"Restart is
required"
message to
alarm history
Semi-critical Discharge
or Standby
Restart is required for
scheduled
maintenance
30 seconds Select Restart Restarts IACS
immediately
Records
"Restart is
required"
message to
alarm history
No action taken
during message
duration
Restarts IACS
Records
"Restart is
required"
message to
alarm history