Alarms
160 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Arrhythmia ranges and defaults
Parameter Alarm priority
default
Rate (default) Count (default) Alarm archive
factory default
ASY High Not adjustable Not adjustable Str/Rec
VF High Not adjustable Not adjustable Str/Rec
VTACH High 100 to 200 (120)
Increments of 10
5 to 15 (10)
Increments of 1
Str/Rec
ARTF Off Not adjustable Not adjustable Off
RUN Medium not adjustable (Rate =
VTACH)
3 to VT count – 1 (3 to
9) changes based on
VTACH
Str/Rec
AIVR Medium Not adjustable =
VTACH rate – 1 (119)
Not adjustable (3) Off
SVT Medium 120 to 200 (150)
Increments of 10
3 to 10 (3)
Increments of 1
Str/Rec
CPT Low Not adjustable Not adjustable Str/Rec
BGM Low Not adjustable Not adjustable Str/Rec
TACH Off 100 to 200 (130)
Increments of 10
5 to 15 (8)
Increments of 1
Off
BRADY Off 30 to 105
(adult 50; pediatric
60)
Increment of 5
Not adjustable
(8)
Off
Pause Off 1 to 3.5 (2.5)
Increments of 0.5
Not adjustable Off