EasyManua.ls Logo

Dräger Infinity C500 - Viewing Alarm Messages

Dräger Infinity C500
610 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
142 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Viewing current alarms
The Current alarms page displays all alarms that
occur during a patient’s monitoring session. The
page is cleared when the patient is discharged.
Each alarm displays the alarm priority, the duration,
and the corresponding message.
To view current alarms
1 Select Alarms... on the main menu bar.
2 Select the Current alarms tab.
Configuring alarm settings temporarily
The following table contains several commonly
used alarm settings that can be configured for a
patient temporarily. These features revert to the
password-protected configuration settings (see
chapter System configuration starting on page 443)
when the patient is discharged.
To configure alarm settings temporarily
1 Select Alarms... on the main menu bar.
2 Select the Settings tab.
Selection Available settings Description
Alarm volume Off, 5%,10 to 100% (in
increments of 10%); default is
50%
NOTE: The 5% setting is only
available when the Minimum
alarm volume setting is set to
5%.
Determines the volume of the alarm tone. You
can never turn the alarm volume lower than the
selected setting for Minimum alarm volume.
NOTES:
Make sure the alarm volume is set so it can be
heard in the monitoring environment.
The setting Off for the alarm volume is only
available under the following circumstances:
When the Cockpit is in OR mode or
assigned to an ICS.
When the Minimum alarm volume feature
is set to Off.
Pulse tone volume Off
5, 10 (default) to 100% (in
increments of 10%)
Determines the volume of the pulse tone.
Attention tone
volume
–Off
5, 10 (default) to 100% (in
increments of 10%)
Determines the volume of the attention tone or
deactivates the attention tone.

Table of Contents