System configuration
470 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Reports setup – Reports setup page
The following table lists the available settings of the
Reports setup page. To access this page, see
page 467.
Biomed setup
This section describes several pages accessible
only to authorized personnel. All Biomed pages are
password protected.
To access the biomed pages
1 Select System setup... on the main menu bar.
2 Select the Biomed tab.
3 Enter the password and select Enter.
4 Select one of the following tabs:
Country
Units of measure
–Patient monitor
Name service
–Network setup (select either the Infinity or
Hospital tab)
Printer setup
Recorder setup
Service
IT setup (select the desired tab such as
Web browser, Innovian, and so on).
NOTE
Reports printed on a laser printer use the settings
in the Reports setup page not the settings de-
fined in the trend setup pages.
Selection Available settings Description
Waveform delay
[s]
6, 10 (default), 15 s Determines the amount of delay (pre-event)
data included in a timed strip report. Delay data
refers to data that originated before the report
was initiated. A marker on the report marks
where the delay data ends and the real-time
data starts.
Waveform
duration [s]
10, 20 s (default) Determines the length of a strip report.
Trend duration
[hr]
1, 2, 4, 8, 12, 24 (default), 48,
72, 96 hr
Determines the graphical trend interval on the
graphical trend report.
Table interval
[min]
1, 5, 10, 15 (default), 30,
60 min
Determines the tabular trend interval on the
tabular trend report.
WARNING
Do not service the Cockpit while monitoring a
patient.