Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 355
Microstream CO
2 monitoring
Calibration Check
The Calibration check tab is used to perform the
following tasks:
Viewing the remaining hours until the next
service check is due for the Microstream
MCable
Viewing the date of the last calibration
Performing a calibration check
A calibration check verifies that the Microstream
MCable is within the acceptable calibration limits. If
successful, the message CO
2 calibration check
successful displays on the Cockpit. For
descriptions of additional message conditions, see
"Calibration and maintenance" on page 547.
Perform a calibration check according to the
healthcare facility’s schedule guidelines.
Required accessories
A gas canister (with a mix of 5% CO2, 21% O2,
balanced N
2).
Sample line
Contact the biomed or service personnel for the
required accessories if needed.
To perform a calibration check
1 Select the Calibration check tab.
2 Follow the instructions as they display on the
Calibration check window.
3 If the calibration check is unsuccessful, see
"Calibration and maintenance" on page 547 or
contact Dräger-authorized service personnel.
WARNING
Risk of inaccurate patient results
A Microstream MCable that is out of
calibration may provide inaccurate results.
If calibration does not take place as
instructed, the Microstream MCable may be
out of calibration.
Ensure the proper calibration of the
Microstream MCable.