Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 369
Scio Monitoring
Change in slope of ascend-
ing limb. Possible absence
of an alveolar plateau.
Obstruction caused by one of the
following:
Partial obstruction in expiratory
limb of breathing circuit
Foreign body in upper airway
Partially kinked or occluded
artificial airway
Herniated endotracheal or
tracheostomy tube cuff
Bronchospasm
Elevated baseline, with
pronounced slope on de-
scending limb
Faulty ventilator circuit valve
Rebreathing (see above)
Description Cause Capnogram