Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 461
System configuration
Alarm setup – configuring the alarm volume and tones
The following table lists the available settings of the
Volume/ Tone page which controls various tone
settings. To access this page, see page 458.
External device
disconnected
alarm control
On (default)
–Off
When this feature is activated, the Cockpit
displays the message External device
disconnected when a device that is connected
to the Cockpit with the device conncectivity
option becomes disconnected.
Selection Available settings Description
WARNING
If you select Follow HR alarm, ASY, and VF
alarms are not reported if the heart rate and
arrhythmia alarm functions are turned off.
Selection Available settings Description
Minimum alarm
volume
–5%,
10% to 100% (in increments
of 10%);
50% (default)
–Off
Determines which alarm volume settings are
available under the Alarm volume button. This
setting does not affect the volume of the
attention or the pulse tone.
Alarm volume Off, 5%,10 to 100% (in
increments of 10%); default is
50%
Determines the volume of the alarm tone. You
can never turn the alarm volume lower than the
selected setting for Minimum alarm volume.
Make sure the alarm volume is set so it can be
heard in the monitoring environment.
The 5% setting is only available when the
Minimum alarm volume setting is set to 5%.
The Off setting is only available under the
following circumstances:
When the Cockpit is in OR mode or
assigned to an ICS.)
When the Minimum alarm volume feature
is set to Off.
Pulse tone volume Off
5, 10 (default) to 100% (in
increments of 10%)
Determines the volume of the pulse tone.