Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 527
Troubleshooting
!!
(Any
SpO2
MCable)
PLS out of range low The parameter
value is replaced
by - - -
The parameter value
is below the
measurement range
of the monitor.
Check the
patient and
treat if
necessary.
Change the
alarm limits.
PLS out of range
high
The parameter
value is replaced
by +++
The parameter value
is above the
measurement range
of the monitor.
!!
(Masimo
rainbow
SET only)
PVI > (alarm limit)
SpHb > (alarm limit)
SpHbv > (alarm limit)
SpMet > (alarm limit)
SpOC > (alarm limit)
Parameter value The parameter value
is above the upper
alarm limits.
Check the
patient and
treat if
necessary.
Change the
alarm limits.
!!
(Any
SpO2
MCable)
SpO2 > (alarm limit)
PLS > (alarm limit)
!!
(Masimo
rainbow
SET only)
SpHb < (alarm limit)
SpHbv < (alarm limit)
PVI < (alarm limit)
SpOC < (alarm limit)
SpMet < (alarm limit)
Parameter value The parameter value
is above/below the set
upper/lower alarm
limits.
The priority changes
to high (!!!) if the SpO2
value falls more than
10% below the lower
limit. This does not
occur when using
SatSeconds alarm
time with the Nellcor
OxiMax MCable.
Check the
patient and
treat if
necessary.
Change the
alarm limits.
(Any
SpO2
MCable)
SpO2 < (alarm limit)
PLS < (alarm limit)
Priority Message Parameter field Problem Solution