Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 533
Troubleshooting
!!
(Any
SpO2
MCable)
SpO2 sensor
unplugged
2)
Sensor
unplugged
Parameter values
are replaced by ***
–SpO2
intermediate cable
or sensor is
unplugged
SpO2 sensor is
unplugged from
Masimo rainbow
SET MCable
Verify that the
cable and the
sensor are
properly
connected.
Check for faulty
sensor.
1)
%0 is a placeholder for the parameter label SpO2.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
3)
In the parameter field the parameter value is replaced by ***
!
(Any
Masimo
MCable)
SpO2 unrecognized
cable
3)
Unrecognized
cable
Parameter values
are replaced by ***
An incompatible cable
is connected.
Connect the
right type of ca-
ble.
Contact
DrägerSer-
vice
®
.
!
(Any
SpO2
MCable)
SpO2 unrecognized
sensor
Unrecognized
sensor
Parameter values
are replaced by ***
An incompatible
Nellcor or Masimo
SET sensor is
connected.
A reusable SpHb
Masimo rainbow
SET sensor is
connected to an
Masimo rainbow
SET MCable that
does not support
SpHb.
Connect the
right type of
sensor.
Contact
DrägerService
®
.
Priority Message Parameter field Problem Solution