Troubleshooting
544 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
! %0 incompatible
sensor
1)
The M540 has detected
that the used
mainstream sensor is
not compatible with the
selected sensor type
setting
(reusable/disposable)
Secretions in the
adapter
Sensor zero drift
High inspiratory CO2
concentration
Use the airway adapter
type the system is
configured for or adjust the
airway adapter setting.
If the message persists,
clean or replace the airway
adapter.
If the message persists
even though the correct
airway adapter type is
selected and the airway
adapter is clean, zero the
sensor.
If the message still persists,
the inspiratory CO
2 value
might not be accurate.
Check the patient and
ventilation.
!! CO
2 out of range high
1)
The parameter signal is
outside the measuring
range of the monitor.
Check the patient and treat
if necessary.
Check the equipment and
replace if necessary.
None CO
2 please zero Instructional message for
the mainstream sensor only.
Zero the mainstream sensor.
! %0 sensor too warm
1)
The CO2 mainstream
sensor is too warm due to
ambient temperature.
Unspecified accuracy at
ambient temperatures
above 40 °C (104 °F).
The sensor will return to
normal operation at
ambient temperatures
below 40 °C (104 °F). If not,
replace the sensor and
contact DrägerService
®
.
! CO
2 MCable
unplugged
1)
The CO2 sensor is
disconnected.
Check the CO2 connections.
! CO
2 MCable failure The CO2 sensor hardware
failed due to a corrupt
EPROM (erasable
programmable read-only
memory) chip.
Contact DrägerService
®
.