Troubleshooting
564 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low %0 value temporarily
unavail.
O2 parameter has unknown
accuracy, possibly due to:
Polluted ambient air
during zeroing.
Electromagnetic distur-
bances.
Overheating.
Ensure clean ambi-
ent air.
Remove radiating de-
vices (e.g., tele-
phone).
Check ambient tem-
perature.
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Wait for auto zeroing
to complete.
Power cycle the gas
analyzer.
Change vaporizer
settings.
Call DrägerService
®
.
Medium RRc > # Respiratory rate has exceed-
ed the limit.
Check ventilation.
Medium RRc < # Respiratory rate is below the
limit.
Check ventilation.
Medium RRc apnea No breathing or ventilation. Start manual ventila-
tion.
Check ventilation set-
tings.
Check spontaneous
breathing ability of
the patient
Sample line is not connected. Connect sample line to
breathing circuit
Medium %0 out of range high RRc has exceeded the upper
limit of the measuring range
for Scio.
Check ventilation.
Low Sample line blocked Sample line or patient-side
filter is occluded.
Check sample line, water
trap, and patient-side
filter.