102 Instructions for use Oxylog 3000 plus SW 1.n
Checking the safety valve
For safety inspection, a flow can be generated to
test the safety valve.
Displaying the technical logbook
Any technical errors and/or special occurrences,
such as activation of a software option, completion
of the device check and device calibration, are
listed in chronological sequence.
The entries in the logbook cannot be deleted and
are retained when the device is switched on and off
or after a power supply failure.
A maximum of 570 logbook entries can be
displayed on the screen. When the capacity of the
logbook is reached, the oldest entries are
overwritten.
Display (example):
To advance to the next page:
Select line Page, confirm and turn the rotary
knob.
Displaying the user logbook
The operational phases (including switching the
device off and on) with the device settings and
the time are listed in chronological order.
The entries in the logbook cannot be deleted and
are retained when the device is switched on and
off or after a power supply failure.
A maximum of 935 logbook entries can be
displayed on the screen. When the capacity of
the logbook is reached, the oldest entries are
overwritten. For alarms, only the occurrence of
the alarm condition is recorded.
To advance to the next page:
Select line Page, confirm and turn the rotary
knob.
Displaying contact details for
maintenance and service
The maintenance schedule and contact details for
maintenance personnel are displayed. The
message Service date overdue ! is shown if
service is needed.