Service
134 Instructions for use Oxylog 3000 plus SW 1.n
Maintenance
Maintenance schedule
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
WARNING
Risk of faulty components
Device failure is possible due to wear or mate-
rial fatigue of the components.
To maintain the proper operation of all compo-
nents, this device must undergo inspection
and preventive maintenance at specified inter-
vals.
WARNING
Risk of electric shock
Before performing any maintenance work, dis-
connect all electrical connectors and gas con-
nectors from power supply and gas supply.
Component Interval Measure Personnel responsible
Dust filter Every 2 years Replace
1)
Service personnel
Internal rechargeable
battery
Every 2 years
When the battery runtime
is no longer reached
2)
Replace Users
Device Every 2 years Inspection and pre-
ventive maintenance
Service personnel
1) The dust filter can be treated as household waste.
2) Refer to "Technical data" section for battery operating time.