Operation
72 Instructions for use Oxylog 3000 plus SW 1.n
Spn-CPAP, Spn-CPAP/PS
SpnCPAP
Spontaneous Continuous Positive Airway
Pressure
For patients with sufficient spontaneous breathing.
If Pressure Support (PS) is not active, the patient's
spontaneous breathing is supported only by an
increased PEEP.
Set the ventilation pattern with the controls below
the display:
Maximum airway pressure Pmax
–O
2 concentration FiO2
The following can be set on the display:
Positive end-expiratory pressure PEEP
Pressure support
Δ
Psupp above PEEP
NIV – non-invasive ventilation
See "Non-invasive ventilation (NIV)" on
page 75.
Insp.term.%PIF
Inspiration termination criteria of pressure
supported breaths as percentage of the peak
inspiratory flow (PIF).
Hose type
The selected hose type must match the hose
type in use. Otherwise a correct volume
measurement cannot be guaranteed.
–CO
2 cuvette type (optional)
15737171
1513717115237171
NOTE
The cuvette windows of the reusable cuvette and
disposable cuvette have different optical proper-
ties. Therefore, the correct cuvette type must be
selected in the settings menu. Otherwise the zero
point will be shifted by as much as ±8 mmHg CO
2.