Instructions for use Oxylog 3000 plus SW 1.n 93
Monitoring
Checking the CO2 filter during
ventilation
1 Remove the CO2 sensor from the cuvette.
2 Connect the CO
2 sensor (A) to the test filter (B).
3 Press the Settings key.
4 On the screen, select and activate the CO
2
filter check – Run line.
5 Confirm. The filter check starts and the line
displays Busy. During the filter check, the
message ! CO2 meas. inaccurate/failed is
displayed.
During the filter check, ventilation settings can
be changed.
After a successful filter check, the message
Pass is briefly displayed.
6 Attach the CO
2 sensor (A) back to the cuvette.
If the check was not successful:
Oxylog 3000 plus displays the alarm message
!!! CO
2 Filter check failed. The measured
CO
2 value is outside the permissible tolerance
range.
Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
CO2 filter check.