Instructions for use Oxylog 3000 plus SW 1.n 55
Getting started
Disposable breathing circuit for adults
1 Make sure that the angled connector (A) is
connected to the flow sensor.
2 Connect the connector (B) of the test lung,
diameter 7 mm, to the angled connector.
The connector simulates the resistance.
3 Connect the bag (C) of the test lung. Refer to
"List of accessories" on page 163.
Switching the device on
1 To switch the device on,
briefly press the key (A).
The device performs a self test and the user is
prompted to activate the configuration menu or
device check:
Press rotary knob for device check and
configuration
2 Press the rotary knob (B) to confirm, before the
progress bar is complete. The start-up screen
appears:
053
NOTE
Oxylog 3000 plus determines MVe and VTe values
assuming BTPS conditions, see page 141.
Consequently, the measured values of a patient
are not identical to those of a test lung with its expi-
ratory gas in the ATPD (Ambient Temperature and
Pressure, Dry) state. Therefore, when a test lung
is connected, the MVe and VTe values indicated
on the display may differ from the MVe and VTe
values set by the user.