Instructions for use Oxylog 3000 plus SW 1.n 39
Assembly
External power supply
External power supply with
DC/DC converter
The DC/DC converter must be used to connect
Oxylog 3000 plus to on-board power supply
systems, e.g., in emergency vehicles. It can be
used with the following voltages: 12 VDC, 24 VDC
or 28 VDC. The on-board power supply must be
protected by a 10 to 16 A DC fuse.
Oxylog 3000 plus cannot use the DC input power
outside of this range.
Mount the DC/DC converter on a flat wall and make
sure the wall is solid enough to support the bracket.
Use all four mounting holes (screw size M4).
1 Plug the large connector (A) of the
DC/DC converter into the on-board supply.
2 Plug the small connector (B) into the
DC socket (C) of Oxylog 3000 plus.
3 When Oxylog 3000 plus is correctly connected
to an external power supply, the indicator
(D) lights up.
WARNING
The DC/DC converter should be used in dry
locations only.
Risk of electric shock and risk of damage to
device.
016