Configuration
98 Instructions for use Oxylog 3000 plus SW 1.n
Setting the start-up settings
The ranges of the settings are: The default settings are:
* Brightness-min: screen brightness level in power
save mode. Refer to "Screen brightness" on
page 80.
Parameter Range of settings
Ventilation mode
Trigger 0 (OFF) 1 to 15 L/min
PEEP 0 to 20 mbar
I:E or Ti Configurable I:E or Ti
I:E 1:100 to 50:1
Ti 0.2 to 10.0 s
Tplat 0 to 50 %
ΔPsupp 0 to 35 mbar
Slope FLAT, MEDIUM, STEEP
Pinsp 3 to 55 mbar
O
2-Flow 0 to 15 L/min
NIV ON, OFF
Tapn 0 (OFF) to 60 s
VTapn 250 to 2000 mL
VTapn (ped.) 50 to 250 mL
RRapn 12 to 60 bpm
RRapn (ped.) 12 to 60 bpm
MVe-high alarm 2.0 to 41 L/min
MVe-low alarm 0.5 to 40 L/min
RR-high 10 to 100/min
HME correction ON, OFF
AutoFlow ON, OFF
Brightness-min* 1/4 to 4/4
Brightness 1/4 to 4/4
Alarm volume 1/4 to 4/4
etCO
2 high alarm 1 to 100 mmHg
etCO
2 low alarm 0 to 100 mmHg
CO
2 unit mmHg, kPa or Vol.%
Hose type Adult disposable
Adult reusable
Pediatric disposable
Cuvette type Disposable
Reusable
Parameter Default setting
Ventilation mode VC-CMV
Trigger 0 L/min at VC-CMV, VC-AC
and 3 L/min at VC-SIMV,
SpnCPAP, PC-BIPAP
PEEP 5 mbar
I:E or Ti I:E
I:E 1.0:1.5
Ti 2.0 s
Tplat 0 %
ΔPsupp 0 mbar
Slope (pressure rise
time)
MEDIUM
Pinsp 20 mbar
O
2-Flow 10 L/min
NIV OFF
Tapn 0 s (when switched on for a
minimum of 15 s)
VTapn 500 mL
VTapn (ped.) 100 mL
RRapn 12/min
RRapn (ped.) 27/min
MVe-high alarm 40.0 L/min
MVe-low alarm 0.5 L/min
RRsp-high alarm 100/min
HME correction OFF
AutoFlow OFF
Brightness-min* 1/4
Brightness 3/4
Alarm volume 3/4
etCO
2 high alarm 50 mmHg
etCO
2 low alarm 33 mmHg
CO
2 unit mmHg
Hose type
Adult disposable
Cuvette type Disposable