Instructions for use Oxylog 3000 plus SW 1.n 149
Technical data
Declaration of hazardous substances in
accordance with Regulation CLP
1272/2008 Annex VI Part 3
Certain materials of this product contain the
following substances in a proportion exceeding
0.1 % by mass:
Lead (CAS No. 7439-92-1)
This product is safe to use for patients who are
sensitive to the indicated substances.
Dräger is aware of the following residual risks:
–None
Device specifications
Sound pressure level of typical ventilation <45 dB (A) at a distance of 1 m.
Sound pressure level of alarm signals 52 to 75 dB (A) at a distance of 1 m.
Sound pressure level of the auxiliary alarm >45 dB (A) at a distance of 1 m.
Dimensions (W × H × D)
Basic unit 294 × 188 × 179 mm (without handle and
protection bar)
Protection bracket 75 mm (in addition to the width of the basic unit)
AC/DC power pack 150 × 37 × 64 mm
DC/DC converter 160 × 35 × 60 mm
Weight
Basic unit without internal
battery
Approx. 5.3 kg
Basic unit with internal
battery
Approx. 5.8 kg
AC/DC power pack Approx. 0.6 kg
DC/DC converter Approx. 0.6 kg
Center of gravity (basic unit)
Width from left 152 mm
Height from bottom 83 mm
Depth from rear 82 mm
Screen
Technology Electroluminescence (EL)
Pixels 240 x 128
Visible area 108 × 56 mm