Technical data
148 Instructions for use Oxylog 3000 plus SW 1.n
DC/DC converter IP classification IP42
Temperature range –20 °C to +50 °C
Input 12/24/28 VDC; 5 A/2.5 A/2.1 A
Output 19 V ±0.5 V/2.6 A
The intended use of the DC/DC converter is in vehi-
cles.
Gas supply
From a central gas supply system or from an
O2 cylinder.
O
2 supply pressure 270 Pa to 600 kPa at 100 L/min
Supply gas Medical oxygen
Connector for O
2 supply either:
NIST
1)
acc. to EN 739/ISO5359 or
DISS
2)
acc. to CGA V5-1989 or
N-F
3)
S90-116
WARNING
Only use medical grade oxygen.
Gas cylinders and pres-
sure reducers
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applica-
ble regulations and have been approved.
Pressure reducer Must have a blow-off valve on the output side to limit
the outlet pressure to approximately 1000 kPa in
the event of a fault.
Gas consumption for
internal control
Average 0.5 L/min
Accuracy of gas
consumption indication
15 % or ±1 L/min, whichever is greater.
1) NIST = Non-Interchangeable Screw-Threaded
2) DISS = Diameter Index Safety Systems
3) N-F = Norme française